Diluter errors
If the display shows dilution errors it indicates that the current sample has not been diluted correctly e.g. -21% Diluent added to the sample. After the sample measurement the dilution rate will be reported as: EDTA 079.
Dilution errors can be caused by:
|
Solution
|
Irregular filling speed due to poor vacuum.
|
Check the vacuum settings.
|
Blocked sample probe.
|
Remove the blockage.
|
Blocked mixing piece
|
Unblock the the mixing piece by using a syringe with hot water.
|
Sample tube pinch valve error.
|
Check if the sample tube pinch valve is working. -> -> ->
|
Sample tube not correct in the sample tube pinch valve.
|
Check the sample tube is still fitted correctly in the pinch valve.
|
Insufficient sample volume.
|
Check before sampling if the sample tube has sufficient blood volume.
|
Wrong diluter settings
|
Check/change diluter settings in software
See Menu selection (diluter) for all settings.
|
Display shows "Diluter failure"
Diluter failures may be caused by:
|
Solution
|
Mechanical obstruction.
|
Clear obstruction.
|
Diluter power cable loose.
|
Put the Power cable back on the connector.
|
A defective diluter motor.
|
Replace diluter motor
|
Top or bottom position sensor failure.
|
Replace the faulty sensor
|
Broken flexible print cable or connector.
|
Replace the print flex cable
Check if the connector on the diluter board is not damaged or broken.
|
Motor tacho failure.
|
Replace the diluter motor.
|
|