Osmoscan options

This is only available when “Enable Osmoscan” is checked.

Settings parameter

Description

Range

Enable Osmoscan

Checkbox, Turns this option on/off

-

Osmoscan shear stress (Pa)

Default = 30

[10, 50] (Pa)

Step size (mOsm/kg)

Default = 10

[1, 10] (mOsm/kg)

Pump speed (ml/s)

Default = 0.053

[0.0053,0.0] (ml/s)

Valid determination per measurement *

Default=10

2-20

Total scan time (s)

Estimated time

540 (s)

Total consumption (ml)

Estimated volume of Osmo LOW and HIGH

(ml)

Calc. area between:

Checkbox, Min or Default = 75

[1,1000]
(mOsm/kg)

…and (mOsm/kg)

Default = 450

[1,1000]
(mOsm/kg)

* This option is available in software version 6.02 and up, the standard setting in older software versions can interfere the results of the Osmoscan