Troubleshooting table
Field
|
Description
|
Check/Remedy
|
Waste container
|
Container indicates waste full when it is not. Waste liquid is clinging to the inside of the bottle and gives a false reading.
|
- Flush bottle thoroughly with hot water or replace the bottle.
|
Power
|
No power
|
- Check the fuses at the main power inlet.
- Check cables
- Replace if necessary.
- Check the fuse on the PC main board
|
Camera
|
No camera image
|
- Check connection cable. Reconnect if necessary.
- Check Camera driver is not defect.
|
Camera
|
Disturbed image
|
|
Fluid flow
|
There is no fluid flow or not enough fluid flow in one or more tubes
|
- Check the liquid levels in the bottles
- Check the functioning of the pinch valve, in case of stitching: perform a Prime all
|
In case of an incorrect image, not sharp and/or black dot is not in the middle, the laser has to be adjusted or even be replaced.
To check the connection on the camera, the handle on the right side has to come off to check if the connection is good.
|