Error Messages (Oxygenscan)

Message

Description

Solution

No PT100 connected!

No PT100 connected to the device or PT100 not correctly connected.

Check the PT100 presence and its soldered connection to the PCB.

No sensor connected!

No sensor is connected (Amplitude < 1000).

Check the optical coupling to the sensor, for example the fiber head2sensor alignment

Signal amplitude too low!

Signal amplitude < 3000.

Check the optical coupling to the sensor, for example the fiber head2sensor alignment

SD card error!

SD card not inserted or defect.

Check the SD card presence and functionality

Reference amplitude out of range!

Contact service team

Device has to be sent to RRM for service

Photo diode saturated!

Ambient light too high or signal strength too high

Send measurement sensor data (Phase, Amplitude Temperature and Calibration data, ideally the repo) to RRM

Reference overflow!

Reference strength too high.

Contact service team. Device has to be sent to RRM for service

Signal overflow!

Signal strength too high.

Decrease signal current.

PME timeout

 

Contact service team. Device has to be sent to RRM for service

Pressure sensor defect!

Pressure sensor out of function.

Contact service team. Device has to be sent to RRM for service

SD card full!

SD card is full.

Backup and wipe SD card

Pulse counter overflow!

The pulse counter has reached the pulse counter overflow limit.

Device has to be sent to RRM for service

Temperature sensor not available!

The temperature sensor was activated but it is not present on this device.

Device has to be sent to RRM for service

Pressure sensor not available!

The pressure sensor was activated but it is not present on this device.

Device has to be sent to RRM for service

Unknown calibration parameter!

Unknown calibration parameter in calibration file.

Wrong calibration file installed or file is corrupted. Check calibration file

Unknown calibration mode!

Unknown calibration mode in calibration file.

Wrong calibration file installed or file is corrupted. Check calibration file

Unknown measurement mode!

Unknown measurement mode in calibration file.

Wrong calibration file installed or file is corrupted. Check calibration file

Unknown 2nd calibration type!

Unknown second calibration type in calibration file.

Check calibration file