Manual home Send Feedback Print

Previous Topic

Next Topic

Book Contents

Book Index

Sample screen

To Sample screen To History screen To reagent screen To Maintenance screen To General settings screen To Service screen Sample Status Sample information Pipette Sample codes Click here for pipette info Pipette belt, double click on pipette for detail information pipette Pipette belt, double click on pipette for detail information pipette Version information Close program

The main menu is displayed during operation. To access other menus, select the required tab on the display.
The following screens are selectable via the associated tabs:

  1. Sample screen
  2. History screen
  3. Reagents screen
  4. Maintenance screen
  5. Settings screen
  6. Service screen

The above picture is an example of the Sample screen of the ESR analyzer unit in the normal operation mode. If the software is set to service mode is this indicated with a yellow bar next to the menu.

Service mode

When the ESR analyzer unit is running in the Service mode all kinds of settings can be changed and the instrument will run with the changed settings. For instance, when ESR time is set to 12 minutes, the Carousel will move according this time setting to be in time at the measure position.
Do not perform normal sampling when Service mode is activated.

When the ESR analyzer unit is running in the NORMAL MODE, the instrument uses the standard saved settings. For instance the ESR time is set back to 60 minutes or 30 minutes according the used method.
The User Manual button is at the bottom of the menu. Click this button to open the Interrliner User manual.

In This Section

Main screen sections

Carousel:

Pipette

Rack:

Sample mode button:

Version information:

Sample information:

Status:

Main screen with keyboard

Pipette information

See Also

Starrsed program

Software version

History screen

Reagents screen

Maintenance screen

Settings screen

Service screen

To top of page Mechatronics home Send Feedback Print