Explain ON and OFF selection
The following selections can be switched ON or OFF:
- ON: ESR's are measured after 30 minutes.
- OFF: ESR's are measured after 60 minutes.
- ON: The dilution rates of all samples are shown in the status line on the screen directly after aspiration.
- OFF: Dilution rates are only shown if they are outside the selected accepted range.
- ON: Samples are presented in EDTA sample tubes. The samples are diluted in the Starrsed.
- OFF: Samples are presented in pre-diluted CITRATE sample tubes. Dilution on the Starrsed is switched OFF.
- ON: A graphical presentation from the measured sample data is shown on the Main screen.
- OFF: Default setting, no graph is shown.
-
- OFF: The sample probe motor will push the sample probe (inner needle) down to the set depth.
- ON: The sample probe motor will stop when a certain current limit is exceeded and returns to the home position.
- Reset the switch from ON to OFF and ON again.
- ON: A temperature corrected value for the ESR is shown besides the actual measured value.
- OFF: Only the actual measured ESR value is shown.
- OFF (Default setting): The carousel is filled with optimum pipette usage but completion of a sample rack takes a little longer. The rotation sequence of the carousel is evenly divided. In 60 minute method, the carousel moves one position every 40 seconds (approx.). In 30 minute method, the carousel moves one position every 20 seconds (approx.).
- ON: The carousel is filled with optimum sample speed. The sample tubes in the rack are faster available for the user but the carousel shows more unused pipettes.
- ON: If keyboard input is required, a virtual keyboard automatically pops up on the screen.
- OFF: No pop-up screen of virtual keyboard.
- This will print the measured result on a single line directly to a dot matrix printer. If this option is switched ON, it is also possible to print a new header at the top of the results.
When other printers then dot matrix are used, every result is printed on one new page.
- ON: It is not possible to run the same sample ID as long as this ID is in the pipette carousel data buffer
- OFF: It is possible to run the same sample ID, even when it is still stored in the carousel data buffer.
Return to General settings (General settings, General settings, General settings)screen.
|