Different Starrsed Control results when comparing instruments

Problem:

Different Starrsed Control results when comparing instruments

Action:

  • Check temperature
  • Check temperature settings
  • Check if results on individual instruments are stable

Additional information:

The results of Starrsed Control are calculated to a result with temperature correction. The combination of measured temperature, the temperature setting and the round-up in the calculation can cause differences in values. Also status and configuration of each individual instrument can cause differences in values.

If results are stable on each instrument individually, sample results can be accepted.

Applicable

Interrliner
Starrsed RS
Starrsed TL
Starrsed ST
Starrsed Compact

FAQ 19007

Related FAQ's

Starrsed Control results are consistently too high

Starrsed Control results are consistently too low

What if Starrsed Control measurement is out of range (FAQ)

What to do if a measurement with Starrsed Control is not exactly the expected value

Normal value is always 6mm with Starrsed Control (FAQ)