Block diagrams over adjusting the fill nozzle

Move the fill nozzle to the pipette home position (See menu selection how to obtain the movement).

For a better grip on the motor place a piece of tubing on the motor axis

pvc_tube_small

To see the correct sequences on the LED 1 and 2 move the fill nozzle motor manual and watch LED 1 comes ON first and then LED 2 comes ON next.

3. Pipette position. The fill nozzle is engaged with the pipette.

4. From pipette to home position. This is the movement from the pipette to the fill nozzle cap.

fill_nozzle_board_sequence_3_small

fill_nozzle_board_sequence_4_small

Detailed Fill nozzle pipette home position

Detailed Fill nozzle to Home position

After the correct adjustment;

  1. Remove the PVC tube from the nozzle motor shaft.
  2. Place the rear panel and reconnect the earth cable, tighten the eight hex.

Return to Led status of the four stages.

See Also

Led status of the four stages

Nozzle unit adjustment Version 1 (ESRI039205)

Help to make the correct sensor adjustment

If the correct setting cannot be reached

To top of page Mechatronics home Send Feedback Print