Diluter start sensor replacement
- Unscrew the hex screw on the ESRI050908 or ESRI050922 mixing piece assembly.
- Carefully disconnect the flat cable from the sensor board.
- Remove the sensor board from the mixing-piece and replace this with a new sensor board.
- Carefully connect the flat cable to the sensor board.
- Screw the hex screw back in the mixing-piece.
- Check the diluter start sensor value. Default value 400-700.
See Sensor information Diluter start sensor
|