Diluter start sensor replacement
  1. Unscrew the hex screw on the ESRI050908 or ESRI050922 mixing piece assembly.
  2. Carefully disconnect the flat cable from the sensor board.
  3. Remove the sensor board from the mixing-piece and replace this with a new sensor board.
  4. Carefully connect the flat cable to the sensor board.
  5. Screw the hex screw back in the mixing-piece.
  6. Check the diluter start sensor value. Default value 400-700.

See Sensor information Diluter start sensor

See Also

Hardware replacement (diluter)

Diluter Motor assembly

Diluter Motor top sensor

Diluter Motor home sensor

Diluter Pulse disk

Diluter EDTA flow sensor with ball bearing

Replacement of the EDTA electronic diluent flow sensor

WI-181 Replace diluter syringe parts