Diluter Motor assembly
- Disconnect the cable (1) from the diluter motor assembly.
- Unscrew the two hex screws from the cover plate and remove the cover plate including the PC Diluter board.
- Loosen the hex screw in the coupling.
- Unscrew the two hex screws which hold the motor base plate ESRI060033 onto the main base plate.
- Pull the motor from the coupling.
- Replace the faulty motor for a new ESRI060930 Motor assembly.
- Push the motor axe into the coupling.
- Assemble the motor base plate ESRI060033 to the main base plate and fasten the two hex screws.
- Tight the hex screw in the coupling.
- Check that the motor turns smoothly.
- Replace cover plate and diluter board and tighten the two mounting hex screws.
- Connect the cable (1) onto the diluter motor assembly.
|