Diluter Motor home sensor

Note: Measure the distance between the home and top sensor and make a note. Normally the distance is 58.5 mm

  1. Unplug the cable with label 4.
  2. Loosen the hex screw that holds the PC sensor board in place.
  3. Remove the faulty sensor board and replace by a new sensor board.
  4. Tighten the hex screw until the sensor board is fixed.
  5. Measure the distance between the two sensors.
  6. Move the new sensor board until the old measured distance is reached.
  7. Tighten the hex screw until the sensor board is fixed.

See Also

Hardware replacement (diluter)

Diluter Motor assembly

Diluter Motor top sensor

Diluter Pulse disk

Diluter start sensor replacement

Diluter EDTA flow sensor with ball bearing

Replacement of the EDTA electronic diluent flow sensor

WI-181 Replace diluter syringe parts