WI-404 Level 4 maintenance

Work instruction Number 404

Page 1 of 4

Purpose:: Level 4 maintenance

Safety: Bio Hazard area

 

Instrument: Starrsed ST

Revision:: 005, 2019

PDF-version

NOTE: Chip card type S is required !

Prepare disinfectant: (if not already prepared).
This disinfectant is for cleaning of all external parts that are exposed to blood.

  1. Wipe the lower tube holder of the needle unit.
  2. Wipe the stainless steel drip plates below the pipette carousel and needle unit.

    drip_plate_needle drip_plate_carousel

  3. In the Main menu use the function [End_of_day wash] .
  4. Check the system for leakage.
  5. Check the waste pump tube condition. Replace the tube if it appears worn or leaking.
    Pump tube
    Pump tube
  6. Inspect sample needle condition:
    Menu Service - Motor control - Needle M31 Press [ ARROW DOWN ] twice to expose the needle tip(s). If the tip(s) are worn, replace the needle.
  7. Check sensors:
    Menu Service - Next - Sensor Status
    Blood sensor: value 110-160
    Measure head: value 40-60

Nozzle O-ring and pipette seal ring replacement

1.

 

Fillnozzle_Oring_01

2.

 

Fillnozzle_Oring_02

3.

Fillnozzle_Oring_03

4.

Fillnozzle_Oring_04

5.

 

FillNozzle_cleaning

6.

 

FillNozzle_Oring_05

5. Clean Fill nozzle with hot water

Note: Replace the flat pipette seal ring BANG030052 and / or replace the nozzle O-ring QWLV050016

7.

FillNozzle_Oring_06

Note: Check if all 5 O-rings QWLV050003 are in place and / or if applicable replace the O-rings QWLV050003

8.

 

FillNozzle_Oring_07

9.

 

FillNozzle_Oring_08

10.

 

FillNozzle_Oring_09