WI-406 Weekly maintenance
Work instruction Number 406
|
Page 1 of 1
|
Purpose: Weekly maintenance
|
Safety: Bio Hazard area
|
|
Instrument: Starrsed ST
|
Revision: 003, 2019
|
PDF-version
NOTE: Chip card type S is required !
Prepare disinfectant: (if not already prepared). This disinfectant is for cleaning of all external parts that are exposed to blood.
- Wipe the lower tube holder of the needle unit.
- Wipe the stainless steel drip plates below the pipette carousel and needle unit.
- In the Main menu use the function [] .
- Check the system for leakage.
- Check sensors:
Menu Service - Next - Sensor Status Blood sensor: value 110-160 Measure head: value 40-60
|