WI-406 Weekly maintenance

Work instruction Number 406

Page 1 of 1

Purpose: Weekly maintenance

Safety: Bio Hazard area

 

Instrument: Starrsed ST

Revision: 003, 2019

PDF-version

NOTE: Chip card type S is required !

Prepare disinfectant: (if not already prepared).
This disinfectant is for cleaning of all external parts that are exposed to blood.

  1. Wipe the lower tube holder of the needle unit.
  2. Wipe the stainless steel drip plates below the pipette carousel and needle unit.

    drip_plate_needle drip_plate_carousel

  3. In the Main menu use the function [End_of_day wash] .
  4. Check the system for leakage.
  5. Check sensors:
    Menu Service - Next - Sensor Status
    Blood sensor: value 110-160
    Measure head: value 40-60