Manual home Send Feedback Print

Previous Topic

Next Topic

Book Contents

Book Index

First Start-up

Steady state

After switching on, the Lorrca Maxsis needs to warm up.

Osmoscan: 5 minutes

Oxygenscan: 1 hour (see also pO2 calibration chapter)

It is recommended to keep the instrument steady state within series of experiments.

Before the instrument can be operated, you have to carry out the following checks and procedures:

  • Check that there are no strange objects (tools, for example) in the instrument.

  • Switch ON the instrument.
  • Start-up the PC.
  • Choose LoRRca software program.
  • Select <Hardware check>.
  • Move the Bob down.

main_screen320x238

  • Choose <Camera>
  • Set camera gain at 250.
  • Close the hood.

HDW

  • Check the image:
    • OK: The instrument is now ready.

camera_alligned

  • Not OK: Misaligned
  • Check position of the projection screen.

camera_misalligned

Alignment of the projection screen:

  • Use the hexagon key 2mm to loosen the projection screen holder.

alignment_projection_screen

  • Align the projection screen until the black dot on the screen is aligned with the laser spot. Note: The projection screen holder itself can also be adjusted (equipped with slotted holes)
  • Tighten the screws of the projection screen holder.

camera_alligned

camera_no_picture

  • Choose <LoRRca IO><Rinse cup> and confirm.
    The tubes are being filled.

Dry BOB and CUP. Do not scratch the cup and do not touch the glass with your fingers to prevent staining! Pay attention to the hole in the BOB.

hwc

After this procedure, the instrument is in the start position for Osmoscan, for Oxygenscan additional steps are required, see Hardware pO2.

See Also

Installation

Pre-installation requirements

Transport

Environmental conditions and setting up the instrument

Checking Lorrca Maxsis functionality

To top of page Mechatronics home Send Feedback Print