Display error history

To Prime/clean screen To check sensors screen To display error history To maintenance history To maintenance info To Sample screen To History screen To reagent screen To Maintenance screen To General settings screen To Service screen Save error history file as

When there is a sensor alarm, an alarm indicator is shown in the tab Maintenance.

All errors that occurred during operation are logged automatically
This list can be used by field engineers to check the status of the instrument and locate possible problems.

This log can be saved e.g. to a memory stick by clicking button Save As ...

Note: since software version 6.12.5 and up is a sample counter included in every error message, e.g. "2020-12-02 16:49:22; 12345; 67890 E9.." with 12345=number of aspirations with successful measurements, 67890=number of total aspirations, including failed measurements.

See Also

Maintenance screen

Prime / Clean

Check sensors

Display maintenance history

Maintenance info

Close

End-of-day-wash options

End-of-day-wash schedule settings