Check sensors

To Prime/clean screen To check sensors screen To display error history To maintenance history To maintenance info To Sample screen To History screen To reagent screen To Maintenance screen To General settings screen To Service screen To explain sensors

When there is a sensor alarm, an alarm indicator is shown in the tab Maintenance.

All functions to check the status of the sensors are grouped under button Check sensors.

  • Check Fill stop sensor: Click the Check button, the OK sign is shown if the sensor value is in range.
  • Check temperature sensor: Value must be equal to the actual room temperature near the pipette belt.
    The value can be set in tab Settings.
  • Check Diluter start sensor: This sensor is only used in EDTA mode. If the diluter does not start during the aspiration, the status of this sensor must be checked.
    Click the Check button, the OK sign is shown if the sensor value is in range.
  • Check Diluent flow sensor: This sensor is only used in EDTA mode. When activated, the indicator Stand-by is activated and the indicator Flow is not. When the button Test is clicked, the indicator Flow must be shown as "activated". After finishing the test, both indicators must be shown as "activated".
  • Check Separator sensor: Click the Check button, the OK sign is shown if the sensor value is in range.
  • Check Flow sensor: Click the Check button, the OK sign is shown if the sensor value is in range.
  • Check Measure sensor:
    Clean sensor first before executing this function.
    Click the Check button, the OK sign is shown if the sensor value is in range.
    Press the button Measure. The pipette currently at the measure position will be measured.
    The results are displayed in graphical form:

take out pipette bza

Measure head start position correct

take out pipette bza

Measure head start position wrong

When a test pipette is installed at the measuring position the result of the test pipette is displayed in the field "ESR (mm)".

Loep

When the sensor is out of range and a red "X" is shown, the sensor values can be checked by turning on the service mode.

 