If it takes too long for the waste pump to empty the waste separator, the system generates a separator error.

Separator error may be caused by:

 

Extensive foam build-up in the waste separator.

Check the separator assembly and connections for possible air leaks.

Waste-tube between waste separator and waste pump is blocked.

Replace the tube.

Waste-tube between waste pump and waste container blocked.

Replace the tube.

Waste pump failure.

Exchange the waste pump cassette. If the error returns, call for service.

Electrical bridge between the waste-level electrodes.

Clean waste separator, see Cleaning waste separator