Errors 165-171 and 210-214

These error messages contain the words "communication" or "response".

  1. Check that the rack transport unit is switched ON.
  2. Clear the error by pressing Clear error button.
  3. When the error occurs again:
  4. Close the Starrsed software.
  5. Switch the ESR analyzer unit OFF.
  6. Switch the rack transport unit OFF.
  7. Switch the ESR analyzer unit ON.
  8. Start the Starrsed software.
  9. Switch the rack transport unit ON.

When the error occurs again, switch OFF all units and call for service.