Errors 183 and 193

Error message: "E193: Elevator motor timeout!" or "E183: Indexer motor timeout!".

  1. Clear the error by pressing Clear error button.
  2. Check resettable fuses at the back. When the fuse was triggered:
    1. Switch OFF the ESR analyzer unit.
    2. Close the Starrsed software.
    3. Check for mechanical obstructions of the elevator/indexer.
    4. Reset the fuse.
    5. Switch ON the ESR analyzer unit.
    6. Restart the PC software.
  3. When the error occurs again, switch OFF all units and call for service.