Flushing liquids

After each sample aspiration the entire system is washed automatically.
If there is no liquid flow:

  • Check that the peristaltic pumps are running. If the pump tubes are worn or leaking, replace the tubes.
  • Check that the pump tubes are installed correctly.
  • Check the tubes between the containers and pumps/valves.
  • Unscrew the cap from the bottle or cubitainer. Check the pick-up tubes in the bottle or cubitainer.
  • Check the tubes for blockages or kinks.

See Also

Troubleshooting

Errors 183 and 193

Errors 165-171 and 210-214

Errors 181-185 and 193-201

Reagents alarm

Separator error

Fill time-out error

Hazy reports

Leaking pipettes

Liquid level sensor not sensing

Air bubbles

Quality control troubleshooting