FAQ&KB home Search Mechatronics home Send Feedback Print

ESR Measurement problems (FAQ)

Question:

How to solve ESR Measurement problems?

Hint:

Could be caused by wrong alignment of the pipettes or in the adjustment of the nozzle.

Answer:

Check if the alignment of the pipettes in the carousel is correct, the pipettes have to make an 90° angle to the plates in the carousel.
For a complete instruction see: Alignment of the pipettes and measure head

Check if the height of the nozzle is correct.
See Nozzle height adjustment.

Applicable for:

Starrsed ST

Related FAQ's

Recurring disabled pipettes (FAQ)

Language setting lost (FAQ)

Citrate mode set-up for Starrsed ST (FAQ)

How many results can be stored ? (FAQ)

Leakage X-Clean after End-Of-Day-Wash (FAQ)

What is the meaning of the numbers on printout during End_of_day_wash? (FAQ)

Error 21 after replacing waste pump tubes (FAQ)

Which Sarstedt tubes can be used on the Starrsed ST? (FAQ)

To top of page Mechatronics home Send Feedback Print