Sensors adjustment (Vacuum and Liquid)

When there is no flow regulator fitted on the compressor itself, there are no sensors adjustments for the overall vacuum and liquid flow needed.

Adjustment for the compressor flow regulator.
650 mbar measured at the separator self, and not less than 600mbar, check adjustment during the wash-all, foaming must be settling under the separator-full sensors.

For adjusting the flow sensor board see section Flow sensor amplifier and position controller.

See Also

Liquid Flow and Vacuum

Quick reference (Flow sensor)

Hardware replacement (Vacuum and Liquid)

Troubleshooting (Vacuum and liquid)