Insert blood sample deformability curve screen

Aspirate Start sampling Aspirate Start sampling

Before inserting blood: Prepare a blood sample for RBC test.

  • Lower the BOB if applicable.
  • Start a new measurement by clicking the <New…(F1)> button in the screen.
  • Enter the measurement ID.

The measurement ID should not contain other characters than digits or letters. It is used to store measurement data. The "Remarks" and "Donation time" fields may be ignored but serve to include sample details. The "Viscosity" field is the viscosity parameter of the test sample, which is mainly determined by the reagent. The indicated value differs per batch and has to be entered at the first run. See chapter Lorrca reagents for obtaining batch information.

Always check whether CUP and BOB are clean and dry before entering a blood sample. When CUP and/or BOB are dirty or some cleaning fluid is left, the measurement results may be influenced.

Depending on the instrument there are two ways that the sample suspension may be inserted.

  • For all instruments by pipetting 880 µl directly into the bottom of the cup, followed by slowly lowering the bob.
  • If there is a bracket with a Luer connection on the bob. Depending on the user preference, a syringe can be used to insert 1600 µl (1,6 ml) of the blood suspension

  1. Check if the suspension level is high enough and without air bubbles.
  2. Close the hood.
  3. Press <OK>

See Also

Performing Deformability curve test

Preparing a blood sample

Adjust video camera deformability screen

Deformability curve result screen

Start new measurement