Gradient pump tube replacement
After 500 measurements/ half year.
![070901 Disconnect tubing](070901_disconnecttubes.jpg)
|
Disconnect water and reagent tubing and pull up the sample needle
|
Lower the BOB and perform two times a function
[ LoRRca IO ] [ Prime all ] [ Start ]
|
![Open drawer](070901_opendrawer.jpg)
|
Open the pump drawer and pull the drawer out.
|
![release_tube_cassette](release_tube_cassette_small.jpg)
|
![replace_tube_cassette](replace_tube_cassette_small.jpg)
|
Release the tube cassette.
|
Replace the tube assembly with new tube assembly following Install LORC070901 tubing (see also Appendix - Parts list spare- and replacement parts)
|
|