Manual home Send Feedback Print

Previous Topic

Next Topic

Book Contents

Book Index

Install LORC070903 tubing set 2

Note: Start to install with the tube end with label 5 which is shown GREEN in the picture.
Next picture: The installed blue tube is now BLUE, continue with the next GREEN tube.

-

LORC070902_tubing_set_1

LORC070902_tubing_set_1

Start with tube end with label 5.

1. Put the short tube into the pinch valve V2 in the back port.

2. Connect tube with "5" on the connector 5.

3. Put the other short tube into the pinch valve V2 at the front port.

4. Connect tube with label 6 on the connector 6.

LORC070902_tubing_set_1

LORC070902_tubing_set_1

5. Take the long end of the tube and run this through the back hole

6. Take connection tube and put the short tube in V1 through the back port.

7. Connect the tube with label 4 to connector 4

LORC070902_tubing_set_1

LORC070902_tubing_set_1

8. Take the long end of the tube and run this through the front hole

9. Connect tube "4" on the bob, connect tube "6" on the EC-sensor

Tubing to waste connector

LORC070902_tubing_set_1

LORC070902_tubing_set_1

  • Connect tube with label 7 on the connector 7
  • Connect tube labelled with waste to the waste bottle

LORC040901

See Also

Tubing replacements

Gradient pump tube replacement

Install LORC070901 tubing

Install LORC070902 tubing set 1

Install LORC070904 tubing set 3

After replacement of all tubing