Manual home Send Feedback Print

Previous Topic

Next Topic

Book Contents

Book Index

Preparing blood sample for RBC aggregation test

A blood sample can be stored at a maximum of one hour (at room temperature). After that the sample cannot be used anymore.

Since aggregation is determined through light back-scatter measurement and indicated as Isc-values are dependent on the amount of oxygenated Hb present, it is important to use fully oxygenated whole blood. When the degree of oxygenation is different in two samples of the same blood, all absolute Isc-values, derived from the Syllectogram, will be different!

  1. Incubate 4 ml of whole blood, with at least 40 ml of air (1:10), in a tube preferably on a roller-bank for 15 minutes, minimum. About 600-900 µl of whole blood is required to perform an aggregation measurement.
  2. Click on Aggregation on the Main screen. The RBC Aggregation start screen appears:
    Rinse Settings Main menu Open New..(F1) Disaggregation level Syllectogram, following cup cessation Iteration curve (I(y )) retrieving the threshold shear rate Patient and sample data Parameters Results area
  3. Press <New...(F1)>

See Also

Performing RBC Aggregation test

Insert blood sample RBC aggregation screen

Adjust temperature RBC aggregation screen

RBC aggregation result screen

Start new measurement

To top of page Mechatronics home Send Feedback Print