Fill and Clean screen

Fill & Clean:
Automatic fill and clean function, each individual pipette on pipette belt will be filled with cleaning solution. During prolonged use of the instrument, proteins are building up in the Westergren pipettes which need to be removed using a strong cleaning agent.
This cycle takes about 90 minutes.
The Fill & Clean function is part of the monthly maintenance procedure.
A warning is shown on the display: <Pipette data will be lost!>.

Select key for using the fill and clean adapter or using  a separeted container with a silicon conneting tube Ok key Return to prime and clean menu

 

By toggling the switch ON the Fill and clean adapter is used.

By toggling the switch OFF the Fill and clean without adapter is used.

See chapter Maintenance Fill and clean procedure for more information.

See Also

Prime / Clean

Priming functions overview

Cleaning functions overview

End-of-day-wash procedure