Fill and clean procedure

Note: Each pipette on the pipette belt will be filled with Starrsed Cleaning Agent, after one hour the first pipette is washed and dried.

Fill and clean with adapter:

Starrsed Cleaning Agent preparation Interrliner ESR analyzer unit: Fill and clean:
This cycle takes about 90 minutes.

  1. Fill the clean adapter EHST110907 with hot de-ionized water. (+/- 150 ml, 80 °C)
  2. Add 15 ml Starrsed Cleaning Agent (QRR 010905) to the hot water in the adapter.
  3. Place the cap on the adapter and mix well.
  4. Put the adapter with cleaning solution on the lower tube holder.
  5. Select Maintenance tab, Prime/Clean, button Fill and clean.

 

fill-clean-ehst-container

Select key for using the fill and clean adapter or using  a separeted container with a silicon conneting tube Ok key Return to prime and clean menu

 

Start Fill and clean procedure:

  1. Select button OK.
  2. The needle goes down and the process is started.
  3. When all the pipettes are filled, the needle goes back to the home position.
  4. After the Fill & clean process is finished, a notification is given to stop the procedure.

    Select key for using the fill and clean adapter or using  a separeted container with a silicon conneting tube Ok key Return to prime and clean menu

     

  5. Remove the adapter from the needle assembly and press "Stop".

The instrument is ready for normal operation.

Fill and clean without adapter:

Select key for using the fill and clean adapter or using  a separated container with a silicon conneting tube Ok key Return to prime and clean menu

Select key for using the fill and clean adapter or using  a separated container with a silicon conneting tube Continue Return to prime and clean menu

  1. The needle comes down.
  2. Fill a container with 150 ml hot (80 °C) de-ionized water.
  3. Add 15 ml cleaning agent. (QRR 010905)
  4. Stir the prepared solution.
  5. Put the container close to the needle location.
  6. Push the silicon tube over the sample probe.
  7. Press Continue.
  8. The fill and clean process is started.