Pump tube replacement
New rinse pump tube assembly ESRI090902.
|
|
New saline pump tube assembly ESRI090903
|
New tube replacement:
- Open left cover.
- Pull pump tube slightly downwards and at the same time towards the front of the unit to release the tube out of the pump plate holder.
- Remove the old tube from the peristaltic pump rotor.
- Disconnect the tubing at both ends of the tube connectors.
- Connect new tubing to both ends of the connectors.
- Place one end of the tube in the pump plate holder.
- Pull the new tube over the peristaltic pump rotor.
- Pull pump tube slightly downwards and at the same time towards the back of the Interrliner.
If the tube is not fitted correctly or is worn the following symptoms can occur.
- Liquid flowing back into the container.
- First glass tube on the pipette belt is not washed sufficiently.
Note: The wider bore tube is for the rinse pump. The narrower bore tube is for the saline pump.
|