Valve control

To Sample screen To History screen To reagent screen To Maintenance screen To General settings screen To Service screen To Elevator/Indexer To LIMS Settings To ESR Unit Settings To Manual control To Service errors To Advanced To Pump control To Needle&nozzle control To Valves&Actuators control Explain Valve control Explain Actuator control

Valves

  1. Valve outer needle: Energizing the outer needle solenoid valve.
  2. Sample control solenoid: Sample control solenoid fill sequence energized. The function of this solenoid is to build up a vacuum in the Westergren pipette before the aspiration starts.
  3. V-Valve wash section: Pipette wash vacuum control valve, controls the main vacuum line between the wash-station and separator.
  4. V-Valve fill section: Vacuum control fill-nozzle / sample probe, controls the main vacuum line between the fill nozzle cap and separator.

Actuators

  1. Rinse actuator: Rinse solenoid active, rinse valve-block down.
  2. Fill actuator: Fill solenoid active, fill valve-block down.

Return to Valve control (Valve control, Valve control, Valve control) screen..