Wash all pipettes:
Before executing this function, check carefully if there are samples in the pipette belt that need to be measured.
Any remaining samples will be washed away and will NOT be measured!
All pipettes on the pipette belt are washed and dried once.
A warning is shown on the display: <Pipette data will be lost!>.

Prime and clean screen Prime all units Prime De-ionised water Prime Disinfectant Prime Diluent Prime Saline Prime / Clean Check sensor Display enter history Display Maintenance history Program is closed via the  shut down procedure Prime Rinse solution Maintenance screen To History screen To Service screen To Reagents status screen To Setting screen To Sample screen Wash each pipette Wash only sample pipettes Wash all pipettes Fill and clean Shut down procedure

All maintenance functions for the fluid system are grouped under button Prime / Clean.

Prime:

After each reagent change, the fluid system must be primed to fill the relevant tubes with reagent and remove air. This is also part of the daily start-up. Use the applicable button to perform the automatic priming cycle for this reagent:

Prime Rinse solution:
After each measurement, the pipettes are washed and dried automatically.Prime Saline:
After each aspiration, the outer needle, sample probe and fill nozzle are washed with saline.Prime Diluent:
The Diluter prime cycle is 5 strokes of the syringe.Prime De-ionized water:
After each aspiration, the fill nozzle is flushed with de-ionized water.Prime Disinfectant:
During a pipette rinse cycle, a small amount of disinfectant is flushed around the bottom of the pipette and into the waste system.When the Starrsed has been idle for more than eight hours, there might be some backflow of reagents due to gravity. Prime all tubing before sampling by using the function:

  • Prime all units
    All priming functions are sequentially performed one time.

Clean:

  • Wash each pipette:
    When the pipette belt turns one position, the pipette at the rinse position will be rinsed and dried, regardless if it was filled or not.
  • Wash only sample pipettes:
    All pipettes which are currently holding samples are washed and dried once.
    A warning is shown on the display: <Pipette data will be lost!>.

    NOTE: Before executing this function, check carefully if there are samples in the pipette belt that need to be measured.
    Any remaining samples will be washed away and will NOT be measured!
  • Wash all pipettes:
    Before executing this function, check carefully if there are samples in the pipette belt that need to be measured.
    Any remaining samples will be washed away and will NOT be measured!
    All pipettes on the pipette belt are washed and dried once.
    A warning is shown on the display: <Pipette data will be lost!>.
    Fill & Clean:
    This button starts the Fill & Clean procedure. During prolonged use of the instrument, proteins are building up in the Westergren pipettes which need to be removed using a strong cleaning agent. This function fills all pipettes with a cleaning agent and removes the cleaning agent after a specified time.End-of-day wash:
    All pipettes will be washed once and needle, fill-nozzle and rinse-nozzle (wash station) are primed.

In This Section

Priming functions overview

Prime rinse solution

Prime Saline

Prime Diluent

Prime de-ionized water

Prime Disinfectant

Cleaning functions overview

Wash each pipette

Wash only sample pipettes

Wash all pipettes

Fill and clean screen

End-of-day-wash procedure

See Also

Maintenance screen

Check sensors

Display error history

Display maintenance history

Maintenance info

Close

End-of-day-wash schedule settings

End-of-day-wash options