Troubleshooting / Error messages

Note: Service Troubleshooting is a combination of service troubleshooting and user troubleshooting section from the user manual.

On vacuum adjustment

  • In case of dilution error's and the ADRC (Automatic Dilution Rate Control) is moving to the 60% the yellow orifice what is in the cross checking of flow sensor is blocked
  • In case of dilution error's and the ADRC is moving to the 140% there is a blockage in the needle system.
    • Check the sample probe
    • Mixing-piece

On position adjustment

  • In case of a position error alarm check the rinse position.
  • When not correct set the rinse position,
    • run the quick belt test or
    • run the flow test.

Error messages:

  • Position error
  • Vacuum not stable