Position adjustment

The positioning principle of one pipette on the pipette belt is done with a digital potentiometer.

During the Pipette flow test and Learn carousel position the software is collecting the values of the 84 positions and stores this information in a table. After finishing the test, the stored information is then used to keep track of the 84 pipette positions in the carousel.

Tab Settings Þ Carousel

  • [ Go to Rinse position ]
  • [ Change Rinse position ]
  • [ Learn carousel position ]
  • [ Pipette flow test ]

Note:
Use the function Learn carousel position when the carousel is filled with blood or cleaning agent.
Use the function Pipette flow test only when the carousel is empty

Return to Flow sensor vacuum adjustment (Flow sensor vacuum adjustment, Flow sensor adjustment (KB))