General information (Flow sensor)
The flow sensor amplifier board ES07v03 and ES07v04 (ESRI080907) has two functions aboard the print.
![esri080938_flow_sensor_board](esri080938_flow_sensor_board.jpg)
|
![Position_pot](position_pot.gif)
|
The flow sensor board have one aboard function
- This new version of vacuum differential detection replaces the previous vacuum differential since June 2008.
- Replace the flow sensor board when when the flow test is malfunctioning.
- Replace the digital potmeter when the positioning is malfunctioning.
|
- Digital pot which is used for positioning the pipettes on carousel.
- Replace the digital potmeter ESRI030920 when the positioning is malfunctioning.
|
|