Drive board reference voltage

On the drive board a build-in reference voltage is available. This reference voltage is used for the following sensors:

  • Temperature sensor
  • Fill stop sensor
  • Measure sensor
  • Position pot meters

If the reference voltage needs to be reset to 5.00 Volt be aware that all the above mentioned sensors need to be readjusted to the correct values.

drive_board_5v_pot

Use the earthing pins or earthing pin/metal part of the nearest condensator.

See Also

Sensor information (flow sensor)

Sensor and motor reference (Flow sensor)

Flow sensor vacuum adjustment

Model Flow sensor check using Starrsed main program