Rinse nozzle (wash station) alignment errors

If there is a mechanical obstruction, the rinse nozzle may not align correctly with the pipette. Check that the two pieces of plastic tubing connected to the rinse nozzle have enough slack to allow for movement of the nozzle.
Technical support is needed if the rinse nozzle does not line up with the pipette.