Error
|
|
|
Problem
|
Possible Cause/Solution
|
09
|
E4009 Start pool. IIC comm.fault
|
|
Transport unit received an unknown command from the Start-pool. The command has been ignored by one of the units!
|
- I2C cable(s) not connected properly.
- Bad I2C cable(s).
Note: the boards in the Interrliner are connected in series.
|
10
|
E4010 Transport 1. NO transport unit
|
|
The first Transport unit was not detected by the Start-pool. The Transport-unit did not response!
|
- No power on the Transport board.
- I2C cable(s) not connected.
- Bad I2C cable(s).
Note: the boards in the Interrliner are connected in series.
|
12
|
E4012 Start pool. No front return
|
|
The transport unit was detected but no corresponding Front Return unit
|
- No power on the Front Return board.
- I2C cable(s) not connected.
- Bad I2C cable(s). Note: the boards in the Interrliner are connected in series.
|
13
|
E4013 Start pool. IIC write error
|
|
There was an I2C communication error between the Transport unit and the Start-pool.
|
- No power on one of the boards.
- I2C cable(s) not connected.
- Bad I2C cable(s). Note: the boards in the Interrliner are connected in series.
|
14
|
E4014 Start pool. IIC status error
|
|
There was an I2C communication error between the Transport unit and the Start-pool.
|
|
15
|
E4015 Start pool. IIC read_1 error
|
|
There was an I2C communication error between the Transport unit and the Start-pool.
|
|
16
|
E4016 Start pool. IIC read_2 error
|
|
There was an I2C communication error between the Transport unit and the Start-pool.
|
|
17
|
E4017 Start pool. TOO many Mess./Text
|
|
The Interrliner tried to load all the text messages (according to the selected language), but there were too many text messages in memory.
|
- Corrupted memory. Service engineer can reload the Interrliner software. Note for service engineer: HEX-file can also be corrupted.
- Faulty Flash Memory (...28F010 or ...28F020) on Start-pool board.
|
18
|
E4018 Start pool. TOO many Menu items
|
|
The Interrliner tried to load all the menus (according to the selected language), but there were too many menu items in memory.
|
|
19
|
E4019 Start pool. TOO much Error text
|
|
The Interrliner tried to load all the error text messages (according to the selected language), but there were too many error text messages in memory.
|
|
20
|
E4020 Start pool. NO transport units
|
|
The Start-pool did not detect any of the Transport units.
|
- No power on the Transport board(s).
- I2C cable(s) not connected.
- Bad I2C cable(s). Note: the boards in the Interrliner are connected in series.
|
21
|
E4021 Start pool. Eeprom read error
|
|
There was a communication error between the EEPROM and the main processor during a read operation. Check all settings!
|
- Faulty EEPROM (24C01) on Start-pool board.
- EEPROM not mounted on Start-pool board.
|
22
|
E4022 Start pool. Eeprom write error
|
|
There was a communication error during a write operation. The main processor was not able to store the settings (see below error list) in EEPROM.
|
- Faulty EEPROM (24C01) on Start-pool board.
- EEPROM not mounted on Start-pool board.
|
24
|
E1024 Start pool. Rack slider timeout
|
|
The Slider motor (Start-pool) did not reach the home position within a certain time limit.
|
- Check Timeout time and Speed settings of the Slider motor (Start-pool).
- Slider mechanically blocked.
- Check Slider motor sensor.
- Faulty Slider motor.
- Broken wire(s) to Slider motor.
- Faulty Slider motor driver on Start-pool board.
|
25
|
E2025 Start pool. Rack forward timeout
|
|
The Rack Transport motor (Start-pool) did not reach the front or back position within a certain time limit.
|
- Check Timeout time and Speed settings of the Rack Transport motor (Start-pool).
- Rack Transport mechanically blocked.
- Check transport sensors.
- Check Rack Home sensor (little red switch).
- Faulty transport motor.
- Broken wire(s) to transport motor.
- Faulty transport motor driver on Start-pool board.
|
26
|
E3026 Start pool. Rack emitter timeout
|
|
The Ejector motor (Start-pool) did not reach the left or right position within a certain time limit.
Or the motor was stopped due to a maximum current.
|
- Check Timeout time and Speed settings of the Ejector motor (Start-pool).
- Ejector mechanically blocked.
- Check ejector sensors.
- Faulty ejector motor.
- Broken wire(s) to ejector motor.
- Faulty ejector motor driver on Start-pool board.
|
32
|
E4032 Start pool. No ID match
|
|
The Interrliner transmitted a request to the Host (LIMS), but the answer from the Host contained a different patient ID.
|
- Check Host (LIMS) system.
- Bad cable(s) between Interrliner and Host (LIMS) system.
|
33
|
E4033 Start pool. ESRI timeout
|
|
There was a communication error between the Interrliner and the ESR unit. The communication was interrupted.
|
- Switching off the ESR Unit could cause this error. This is normal.
- Check RS232 cable between Elevator board and Interrliner Transport board.
- If there are I2C errors in the ESR unit or Interrliner: check the I2C cable(s).
|
37
|
W5037 Start pool. ESRI ack timeout
|
|
There was a communication error between the Interrliner and the ESR unit.
The Interrliner did not receive an answer from the ESR Unit after sending a patient confirmation to the ESR unit.
The communication was interrupted.
|
- Switching off the ESR unit could cause this error. This is normal.
- Check RS232 cable between Elevator board and InteRRliner Transport board.
- If there are I2C errors in the ESR unit or Interrliner: check the I2C cable(s).
|
38
|
E2038 Start pool. Startpool rack blocked.
|
|
A rack was detected and transported in the Start-pool. The Rack Transport unit (Start-pool) reached the front position, but the Rack Home sensor did not detect the rack.
|
- Check Start-pool for misplaced rack(s).
- Check Rack Home sensor (little red switch).
- Check Start-pool cross sensor (transmitter & receiver).
- Check Rack transport home sensor.
|
40
|
W7040 Start pool. Startpool full!!
|
|
A new rack is ready to be slide in the Start-pool, but the Start-pool is already full.
This is a text message (no alarm) and will clear itself.
|
- If this message comes up and the Start-pool is not full, check the Start-pool Full sensor.
|
42
|
W5042 Start pool. Rack slider pos.err
|
|
A rack on the Start-pool conveyer belt (Pool-in) did not reach the Rack-in sensor within a certain time limit.
|
- Check Slider unit (Start-pool).
- Check Slider motor sensor.
|
43
|
W5043 Start pool. Feed belt timeout
|
|
A rack on the Start-pool conveyer belt (Pool-in) did not reach the Rack-in sensor within a certain time limit.
|
- Check Timeout time and Speed settings of the Pool-in motor (Start-pool).
- Rack on conveyer belt (Start-pool) mechanically blocked.
- Check Rack-in sensor.
- Faulty belt motor.
- Broken wire(s) to belt motor.
- Faulty belt motor driver on Start-pool board.
|
44
|
W5044 Start pool. Eeprom chksum error
|
|
After reading the settings from EEPROM the calculated checksum did not match with the checksum stored in EEPROM. Default settings are loaded!
Check all settings (see below error list)!
|
- Faulty EEPROM (24C01) on Start-pool board.
|
45
|
W5045 Start pool. Warning: pool empty
|
|
The Rack Home switch (Start-pool) was activated during a rack transportation, but the Start-pool cross sensor detected NO rack.
|
- Removing the rack during transportation and pressing the Rack Home switch by hand could cause this warning. This is normal!
- Check Start-pool for misplaced rack(s).
- Check Rack Home switch (little red switch).
- Check Start-pool cross sensor (transmitter & receiver).
|
49
|
W5049 Start pool. NO host response_1
|
|
The Interrliner did not receive an answer (Ack / Nack) from the Host (LIMS) system after sending an inquiry to the Host.
|
- Check Host (LIMS) system.
- Check protocol settings.
- Check baudrate settings.
- Bad cable(s) between Interrliner and Host (LIMS) system.
|
50
|
W5050 Start pool. NO host response_2
|
|
The Interrliner did not receive the “patient order information record” from the Host (LIMS) system within a certain time limit.
|
- Check Host (LIMS) system.
- Bad cable(s) between Interrliner and Host (LIMS) system.
|
51
|
W5051 Start pool. Error info length
|
|
The Interrliner received an incorrect “patient order information record” from the Host (LIMS) system.
The length of the record was incorrect!
|
- Check Host (LIMS) system.
- Check protocol settings.
- Bad cable(s) between Interrliner and Host (LIMS) system.
|
52
|
W5052 Start pool. NO host response_3
|
|
The communication was interrupted while receiving a “patient order information record” from the Host (LIMS) system.
|
- Check Host (LIMS) system.
- Bad cable(s) between Interrliner and Host (LIMS) system.
|
53
|
W5053 Start pool. NO host response_4
|
|
The Interrliner did not receive an answer (Ack / Nack) from the Host (LIMS) system after sending a “sample result record” to the Host.
|
- Check Host (LIMS) system.
- Check protocol settings.
- Check baudrate settings.
- Bad cable(s) between Interrliner and Host (LIMS) system.
|
54
|
W5053 Start pool. NO host response_5
|
|
The Interrliner did not receive an answer (Ack / Nack) from the Host (LIMS) system after sending a “flag record” to the Host.
|
see Error 53
|
55
|
W5055 Start pool. Eeprom2 chksum error
|
|
After reading the settings from EEPROM the calculated checksum did not match with the checksum stored in EEPROM.
Default Eeprom2 settings are loaded!
Check Eeprom2 settings (see Default settings)
|
- Faulty EEPROM in processor. Change and save settings. Replace processor if this error recurs.
|
56
|
W5056 Start pool. HST fault !
|
|
Warning: an error occurred during rack handling between Start-pool and external device (HST front return line).
Possible jammed rack.
|
- Remove rack and reset external device (HST). Warning is automatically cleared when external device is ready.
- Reset Interrliner if warning is not automatically cleared.
|
62
|
E0062 Transport 1..4. Supply rack lost!
|
|
A rack was lost during transportation from Start-pool to ESR analyser unit.
|
- Removing the rack during transportation could cause this error. This is normal!
- Rack mechanically blocked.
- Check Timeout time and Speed settings of the Transport motor (Transport unit) or the Ejector motor (Start-pool).
- Check Rack Catch sensor (Transport unit).
- Check main conveyer belt motor (Transport unit).
- Check Ejector motor (Start-pool).
|
63
|
E0063 Transport 1..4. ESRI timeout_1
|
|
The ESR analyser unit did not open or close the indexer within a certain time limit.
|
- Check the indexer (clamp). The indexer (clamp) can be tested in menu Indexer.
- Check RS232 cable between Elevator board and Transport board.
|
65
|
E0065 Transport 1..4. ESRI timeout_3
|
|
A rack has been sampled and was ready to be pushed (switch-out) on the conveyer belt, but the Compact did not open the indexer within a certain time limit.
|
- Check the indexer (clamp). The indexer (clamp) can be tested in menu Indexer.
- Check RS232 cable between Elevator board and Transport board.
|
66
|
E0066 Transport 1..4. ESRI timeout_4
|
|
A rack has been sampled and pushed (switch-out) on the conveyer belt, but the ESR analyser unit did not close the indexer within a certain time limit.
|
- Check the indexer (clamp). The indexer (clamp) can be tested.
- Check RS232 cable between Elevator board and Transport board.
|
67
|
E0067 Transport 1..4. Discharge rack lost!
|
|
A rack was lost during transportation.
|
- Removing the rack during transportation could cause this error. This is normal!
- Rack mechanically blocked.
- Check Timeout time and Speed settings of the Transport motor (Transport unit) or the Pool-in motor.
- Check Switch-out sensor (Transport unit).
- Check transport unit motor(Conveyer belt)
|
70
|
E1070 Transport 1..4. Switch_in timeout
|
|
The Switch-in motor (Transport unit) did not reach the home position within a certain time limit.
|
- Check Timeout time and Speed settings of the Switch-in motor (Transport unit).
- Switch-in unit mechanically blocked.
- Check Switch-in sensors.
- Faulty Switch-in motor.
- Broken wire(s) to Switch-in motor.
- Faulty Switch-in motor driver on Transport board.
|
71
|
E1071 Transport 1..4. Switch_out pos. err
|
|
The Switch-in motor (Transport unit) could not move to the home position, because the Switch-out motor (Transport unit) was not on position (home).
|
- Check Switch-out unit (Transport unit).
- Check home sensor of Switch-out unit.
|
72
|
E1072 Transport 1..4. Catch timeout
|
|
The Switch-in motor (Transport unit) did not reach the catch or release position within a certain time limit.
|
- Check Timeout time and Speed settings of the Switch-in motor (Transport unit).
- Switch-in unit mechanically blocked.
- Check Switch-in sensors.
- Faulty Switch-in motor.
- Broken wire(s) to Switch-in motor.
- Faulty Switch-in motor driver on Transport board.
|
73
|
E1073 Transport 1..4. Switch_in pos. error
or
E2073 Transport 1..4. Switch_in pos. error
|
|
The Switch-in motor (Transport unit) did not reach the home position within a certain time limit.
or
The Switch-out motor (Transport unit) could not move the release position, because the Switch-in motor (Transport unit) was not on position (home).
|
- Check Timeout time and Speed settings of the Switch-in motor (Transport unit).
- Switch-in unit mechanically blocked.
- Check Switch-in sensors.
- Faulty Switch-in motor.
- Broken wire(s) to Switch-in motor.
- Faulty Switch-in motor driver on Transport board.
|
74
|
E1074 Transport 1..4. Switch_in out pos.
|
|
The Switch-in motor tried to move from the release to the home position, but the Switch-in motor was not on the release position.
The position is unknown!
|
- Check Switch-in sensors (Transport unit).
|
75
|
E2075 Transport 1..4. Switch_out timeout
|
|
The Switch-out motor (Transport unit) did not reach the home or release position within a certain time limit.
|
- Check Timeout time and Speed settings of the Switch-out motor (Transport unit).
- Switch-out unit mechanically blocked.
- Check Switch-out sensors.
- Faulty Switch-out motor.
- Broken wire(s) to Switch-out motor.
- Faulty Switch-out motor driver on Transport board.
|
90
|
E4090 Transport 1..4 Front ret init error
|
|
Fatal error during initalisation of the Front Return unit.
The Transfer motor did not reach the home or return position within a certain time limit.
|
- Check transfer motor settings.
- Rack mechanically blocked.
- Transfer motor mechanically blocked.
- Check Transfer motor sensors.
- Faulty Transfer motor.
- Broken wire(s) to Transfer motor.
- Faulty Transfer motor driver on Front Return board.
|
91
|
E4091 Transport 1..4 Front ret rack lost
|
|
A rack was lost during transportation on the Front Return conveyer belts.
|
- Removing a rack during transportation could cause this error.
- Rack mechanically blocked.
- Check Timeout time and Speed settings of the Long belt motor (Front Return unit(s)).
- Faulty sensors.
- Faulty Conveyer belt motor.
- Broken wire(s) to Conveyer belt motor.
- Faulty Conveyer belt motor driver on Front Return board.
|
92
|
W4092 Transport 1..4 Transfer motor error.
|
|
The Transfer motor did not reach the home, catch or return position within a certain time limit.
|
- Check transfer motor settings.
- Rack mechanically blocked.
- Transfer motor mechanically blocked.
- Check Transfer motor sensors.
- Faulty Transfer motor.
- Broken wire(s) to Transfer motor.
- Faulty transfer motor driver on Front Return board.
|
93
|
W5093 Transport 1..4 Misplaced rack
|
|
A rack was placed on the wrong side of the Transfer motor.
|
- Remove rack. (warning is automatically cleared when rack is removed)
|
94
|
W5094 Transport 1..4 Warning FR Catch
|
|
The Transfer motor did not catch a rack within a certain time limit.
|
- Check Transfer motor settings.
- Rack mechanically blocked.
- Check front return cross sensor.
|