Diluter is not starting (FAQ)

Symptom :

Diluter system is not starting

Hint :

  1. Blockage in mixing piece
    Check and clean mixing piece
  2. Blockage in sample tube
    Clean or replace sample tube
  3. Aspiration is too slow
    Check flow settings and vacuum
  4. Bad vacuum flow
    Check vacuum system.
  5. Fill vacuum valve error
    Check fill vacuum pinch valve (Correct position, O-ring, wiring)

Additional information :

Troubleshooting diluter not started

Applicable for:

Interrliner
Starrsed RS
Starrsed TL
Starrsed Compact