Symptom :
|
On the printout EDTA value is 001 (EDTA001)
|
Hint :
|
Diluter is not started
|
Action:
|
- Check if the vacuum is present.
- Check the values with the Flow sensor:
Default: Flow 0925..0980..1000 , Abs 320 ±10, Leakage 0045..0050..0055. - Check the value of Diluter start sensor.
Default 400-550-700. - Faulty connection cable or a broken start sensor.
- Pinch valve which holds the sample tube has a bad contact or broken wire at the solenoid body.
|
Additional information :
|
The diluter gets the information from the Flow sensor board and the start information from the start sensor on the mixing-piece.
|
|
Explain check sensor screen
|