On the printout EDTA value is 001 (FAQ)

Symptom :

On the printout EDTA value is 001 (EDTA001)

Hint :

Diluter is not started

Action:

  1. Check if the vacuum is present.
  2. Check the values with the Flow sensor:
    Default: Flow 0925..0980..1000 , Abs 320 ±10, Leakage 0045..0050..0055.
  3. Check the value of Diluter start sensor.
    Default 400-550-700.
  4. Faulty connection cable or a broken start sensor.
  5. Pinch valve which holds the sample tube has a bad contact or broken wire at the solenoid body.

 

Additional information :

The diluter gets the information from the Flow sensor board and the start information from the start sensor on the mixing-piece.

 

Explain check sensor screen

Applicable

Interrliner
Starrsed RS
Starrsed TL
Starrsed Compact

FAQ 30052