Manual home Send Feedback Print

Previous Topic

Next Topic

Book Contents

Book Index

Indexer

Time out error Indexer.

  1. Clear time out error by pressing Clear error button.
  2. Check resettable fuses at the back. When the fuse was triggered:
    1. Switch OFF the ESR analyser unit.
    2. Close the Starrsed software.
    3. Check for mechanical obstructions of the elevator/indexer.
    4. Reset the fuse.
    5. Switch ON the ESR analyser unit.
    6. Restart the PC software.
  3. When the error occurs again, switch OFF all units and call for service.

open_to_resetable_fuse

Resetable_fuss.jpg

See Also

Transport Units

Elevator

Elevator positions

Tube sensor detection test

Tube not in 1 position

WI-206 Barcode adjustment

To top of page Mechatronics home Send Feedback Print