Manual home Send Feedback Print

Previous Topic

Next Topic

Book Contents

Book Index

Nozzle assembly

The nozzle controller board controls the fill nozzle movement, level detection’s, peristaltic pumps, wash station, vacuum valves, vacuum and waste pump.

There are no adjustments necessary on this board. All cable connections to this board has numbers. Make sure when the cables have been removed they are connected again to the correct sockets.

The following mechanical adjustments need to be checked at least once a year.

  • Fill Nozzle and the Rinse nozzle adjustment.

Incorrect adjustment may cause stall during the aspiration sequence.

If the belt has been re-positioned, the wash station alignment needs to be checked.

  • See Menu selection (nozzle) selection how to obtain the movement.
  • The wash station should engage the pipette smoothly.

In This Chapter

Quick reference (nozzle)

Menu selection (ESR unit)

Error messages (nozzle)

Work instructions (nozzle)

Special tools requirement

Block diagrams (nozzle)

Sensor information (nozzle)

Hardware replacement (nozzle)

Hardware adjustment (nozzle)

Troubleshooting / Error messages

Useful information (nozzle)

To top of page Mechatronics home Send Feedback Print