Manual home Send Feedback Print

Previous Topic

Next Topic

Book Contents

Book Index

Explain mixer

The following settings need to be set;

Mixing settings;

  1. Number of turns for long mix.
    • This is the first mix of a new sample rack. The minimum mix value is set default for 8 mixing.
  2. Number of turns short mix.
    • After the Mixing interval time a second mix is started, the number of the second mix is set default for 1 mix.

Check both mixing settings with the Test button. Observe the numbers in the numerical field.

Mixing motor settings;

  1. Mixing direction:
    • Default setting is left.
  2. Mixing speed: default setting 90%
  3. Hold level: default setting 30%
  4. Homing speed: default 20%

After setting use Save settings to save the new settings. If Save settings is not pressed the old settings will be used

See Also

Mixer motor

Explain mixer motor

To top of page Mechatronics home Send Feedback Print