NOTE: Chip card type S is required !
Additional instructions for Level 3:
Replace waste pump tube.
Replace disk filter The disk filter QWLV040001 is the filter which can be found on the waste bottle.
Replace needle sealing block (not applicable for single needle version)
Applicable for Double needle version
|
Be aware of sting danger !!
|
Removal
|
|
|
|
1. Open needle cover to get access to the needle.
|
2. Remove both Knurled screws.
|
|
|
3. Push the needle down (friction possible)
|
4. Place screw in needle sealing block BANG049002.
|
|
|
5. Pull needle sealing block out the needle guide block assembly.
|
6. Remove screw
|
In case of replacement of needle: disconnect tubing
Place new Needle sealing block BANG049002 (and Needle assembly BANG049001 if needed).
In case of needle replacement: connect tubes on the new needle assembly
|
|
1. Stick the needle in the needle sealing block, the needle not completely through.
|
2. Place the needle sealing block into the needle guide block. Push the needle back upwards against the friction.
|
|
|
3. Push needle up till it is lined up to the mounting fix position.
|
4. Place both Knurled screws
|
|
6. After the assembling is finished, the needle need to be set into the correct position in the needle sealing block.
- Insert an empty sample tube into the needle assembly.
- Return to the main screen and go to the sample screen, type a number for the patient ID,
press and press key which stops the sampling. - The needle is now correct located in the needle sealing block.
|
5. Replace Needle cover.
|
|