Quality control procedureStarrsed Control is provided in ready-to-use sample tubes and is used in the same manner as patient samples. Starrsed Control is to be used for the Westergren method with dilution only as prescribed by the "ICSH review of the measurement of the ESR" (2011) and the "CLSI Procedures for the ESR Test; Approved standard; H02-A5" (2011).
For detailed information see the Starrsed Control Package Insert. The contents of one tube of 4.6 ml is sufficient for three Control samples. Do not mix residual material with material from other tubes. Do not re-use empty tubes. See chapter QC Error messages for handling QC errors. If the QC sample is out of range the message "Control out of range!" appears. This warning is also given when patient sampling is started. Press ESC for aborting patient sampling, press ENTER to continue sampling. It depends on the general QC procedure of the Lab if it is allowed to perform patient samples after the warning "Control out of range!".
|
||||||||