QC Error messages

The general ESR errors and warnings are also applied on the QC results, see "ESR Error and Warning code messages"

When the result is within range, no message is shown. When the result is out of range an error message is shown.

Press ESC to continue.

Press ESC to keep this warning active, press Enter to ignore this warning.

Message when the general setting "Temperature Correction" (TCorr) is switched ON:

  • "E141"
    The uncorrected and the corrected result are out of range.

Messages when the general setting "Temperature Correction" (TCorr) is switched OFF:
The software always calculates a temperature corrected result because only temperature corrected results can be compared with the Assay mean value.

  • "E141"
    The uncorrected and the corrected result are out of range.
  • "E142"
    The uncorrected result is out of range, but the corrected result is within range.
  • "E143"
    The uncorrected result is within range, but the corrected result is out of range.

See also Quality control troubleshooting.

The message "Control out of range!" will be shown until a correct QC result is performed or this message is ignored (with continue). When this warning is kept "active"no patient samples can be performed. Depending on the users Quality Control Procedures patient samples can be or cannot be performed after a QC Error message.

See Also

QC Results

QC Result analysis