Sample tubes without barcode
If the sample tube has no barcode or the barcode can not be read, you can enter the sample ID manually.
- Start the .
- Enter the ID number.
- Push the [ START ] key.
|
NOTE: The sampling may not start immediately if a clean pipette is not yet available, but will start automatically when a clean pipette is available within 90 sec.
To prevent the aspiration of a not properly mixed sample, sampling will not start automatically when it takes longer than 90 sec. for a pipette to become available. The operator must then remix the sample and push the [ START ] key again.
|
|