Report 30 Minute mode

 

Column 1

Column 2

Column 3

Column 4

Line 1

ID

 

1234567890

 

Line 2

ESR T. corr

 

75

mm/1h

Line 3

ESR 60:

 

84

mm/1h

Line 4

ESR 30:

 

42

mm/½h

Line A

Aspect

 

Clear

 

Line B

EDTA:

ON

ESR time

30

Line E

15-03-10

13:01

23 ºC

P15

Line D

 

 

 

 

Sample ID=Patient number.

Line 1

ID

 

1234567890

Tc=60-minute ESR result in millimeters, corrected for 18°C. (only in use if temperature correction is active).

Line 2

ESR T. corr

 

75

mm/1H

ESR=Not corrected 60-minute ESR result.

Line 3

ESR 60:

 

84

mm/1h

Hh=Not corrected 30-minute ESR result (only in use if 30 minute mode is active).

Line 4

ESR 30:

 

42

mm/½h

Aspect (clear, hazy).

Line A

Aspect

 

Clear

 

EDTA mode is On or OFF

Line B

EDTA:

ON

ESR time

30

ESR time: Time=Actual sedimentation time in minutes.

Line C

15-03-10

13:01

22 ºC

P15

 

The present date

Present time

Present temperature

Pipette number, in this example pipette 15

Line D is used for error and warning messages and printed in reversed colours.

See Also

Result Printout

Report 60-Minute mode

Analyser ERROR

Results at limit errors

Reporting range

Aspect Hazy