Cleaning

Clean the Instrument once a day. Note the following points:

Never use compressed air for cleaning, as this may cause swarf and fragments to be blown into the instrument.
The use of flammable fluids is not allowed.
Do not use medical alcohol, bleach or any solvent for cleaning, use only DI water or Starrsed X-clean.

 

  • Clean the casing with a soft dust-free cloth moistened with a few drops of subscribed disinfectant.
  • Clean the area around the instrument and remove dust from the floor with the vacuum cleaner.
  • Clean the projection screen with a clean soft cloth or lens paper.
  • Clean the CUP with a soft dust free cloth moistened with DI water or Starrsed X-Clean.