Insert blood sample membrane screen
Before insert blood Preparing a blood sample Stability for RBC test first.
The patient name is not allowed to contain characters other than digits or letters. It is used to store measurement data (if enabled in the Settings menu). The “Remarks” and “Donation time” fields may be ignored but serve to include sample details. The "Viscosity" field is the viscosity parameter of the test sample, which is mainly determined by the suspending PVP medium. The indicated value variates per PVP-batch, see also Settings screen. Always check viscosity in case of performing various measurements on the instrument. Always check whether CUP and BOB are clean and dry before entering a blood sample. When CUP and/or BOB are dirty or some cleaning fluid is left, the measurement results may be influenced.
|
||||||