Mechatronics home Send Feedback Print

Previous Topic

Next Topic

Book Contents

Book Index

Stability curve determination parameters explanation

The following result parameters are given in Results file and printed report:

Parameter

Description

#

Measurement number

Start time

Actual time that was valid after setting a new shear stress and immediately following the image stabilisation time.

End time

Actual time that was valid after the last EI measurement at the given shear stress was performed.

Time (s)

Time, in seconds, since the beginning of the measurement.

Coordinate count

Number of coordinates in the iso-intensity curve that describes the shape of the diffraction pattern.

A (pixels)

Vertical axis of ellipse.

B (pixels)

Horizontal axis of ellipse.

EI

Elongation index

Eof (pixels)

Error of fit. This parameter represents the average distance (in pixel units) of the iso-intensity pixels to the best-fit ellipse that resulted. It reflects how well the iso-intensity coordinates comply with the ellipse model.

See Also

Stability curve determination results screen